ECT Service Department of Biological Psychiatry New York State Psychiatric Institute ECT Device Test and Calibration Form Mecta spECTrum 5000Q Models Testing conducted by: Device Tested: Name Manufacturer/Model Results of Testing (Pass/Fail): Reason for Failure: Date Serial No. COMMENTS I Signature of Authorized Individual Assessing Test Results Form Revision Date: 2/06/01 Date (Page 1 of 15) ECT Device Test and Calibration Form Date Device Type Serial Number-] I I I I Line Isolation Safety Power Cord Resistance (mOhm) Case Leakage, External Lead (ľA) [<500ľAj Power Off Power On -Normal Power On - Reverse Ground Off -Normal_ Ground Off - Reverse Neutral Off - Normal Neutral Off - Reverse Case Leakage, Ground Conductor (ľA) [<58,ľA] Power On -Normal Power On - Reverse 0 (Page 3 of 15) ECT Device Test and Calibration Form Date Device Type Static Impedance Readout Self-test passes in 150-5,000 ohm range, and fails with resistance below 150 ohm and above 5,000 Documentation of Static Impedance Readout Serial Number I Self-Test Range Pass Fail Known Resistor Device Static I pedance Values Low Dose #1 High Dose #1 Low Dose #2 High Dose- #2 150 ohm 200 ohm 300 ohm 400 ohm 500 ohm 600 ohm 700 ohm 800 ohm 900 ohm 1000 ohm 1500 ohm 2000 ohm 2500 ohm 3000 ohm 3500 ohm 4000 ohm 4500 ohm 5000 ohm (Page 4 of 15) I ECT Device Test and Calibration Form Serial Nu C. Date Device Type I Stimulus Control Parameters Warning stimulus beeps three times prior to stimulus delivery/ continuous tone during delivery During continuous tone LCD reads "Stimulus About to Occur", then "Delivering Stimulus." After stimulus reads "Performing Internal Tests" Following delivery, LCD displays treatment parameters (Dym Impedance, Charge, Energy, Current) With release of stimulus control during warning, LCD displays "Treatment Terminated. Stimulus control push button released prematurely. Press clear to proceed" With release of stimulus control during delivery, LCD dispIays "Treatment Terminated. Stimulus control pus button released prematurely. Press clear to proceed" Following stimulus delivery, the chart recorder engages and monitors the installed amplifier 0 channels Pressing Off halts chart monitoring, and reports treatment results Stimulus Warning Pass I Fail Stimulus Delivery Pass I Fail LCD Parameters Pass I F a- ail I Premature Release Pass Fail Premature Release Pass Fail Chart Recorder Engage Pass Fail Treatment Results Pass Fail (Page 5 of 15) ECT Device Test and Calibration Form Date Device Type Serial Number Pulse Width Calibration Low Intensity 20 Hz, .5 sec; .5 A, 200 ohm Independent Obtained Values Voltage + Voltage - Current + Current - Hz 0.2 ms 0.3 ms 0.4 ms 0.5 MS 0.6 ms 0.7 ms 0.8 Ms 0.9 MS 1.0 MS 1.1 Ms 1.2 ms 1.3 ms 1.4 ms 1.5 Ms 1.6 ms 1.7 ms 1.8 Ms 1.9 Ms 2.0 ms 0 (Page 6 of 15) ECT Device Test and Calibration Form Date Device Type Serial Number 0.2 ins 0.3 ms 0.4 ms 0.5 ms 0.6 ms 0.7 ms 0.8 ms 0.9 ms 1.0 ms 1.1 ms 1.2 ms 1.3 ins 1.4 ms 1.5 ms 1.6 ms 1.7 ins 1.8 ms 1.9 ms 2.0 ms I Pulse Width Calibration Low Intensity (Cont.) 20 Hz, .5 sec, .5 A, 200 I -1 SPECTRUM Reported Values current Dynamic Energy Charge Imp Independent Values Current Dynamic Energy Charge Imp (Page 7 of 15) ECT Device Test and Calibration Form Date :1 I Device Type Pulse Width Calibration High Intensity 1 120 Hz, 4 sec; .8 A, 300 ohm I Serial Number I Independent Obtained Values Voltage + Voltage - Current + Currant - Hz 0.2 ms 0.3 ms 0.4 ms 0.5 ms 0.6 ms 0.7 ms 0.8 ms 0.9 Ms 1.0 Ms 1.1 MS 1.2 ms 1.3 ms 1.4 ms 1.5 ms 1.6 ms 1.7 ms 1.8 ms 1.9 Ms 2.0 ms I (Page 8 of 15) ECT Device Test and Calibration Form Date Device Type 2 ms 0.3 ms 0.4 ms 0.5 ms 0.6 ms 0.7 ms 0.8 ms Serial Number Pulse Width Calibration High Intensity (Cont.) 120 Hz, 4 sec; .8 A, 300 ohm SPECTRUM Reported Values Independent Values Current Dynamic Energy Charge Current Dynamic Energy Charge Imp Imp 1.1 ms 1.2 ms 1.3 ms 1.4 ms 1.5 ms s Charge 1.6 ms 1.7 ms 1.8 ms 1.9 ms 2.0 ms (Page 9 of 15) ECT Device Test and Calibration Form Dale Device Type Pulse Frequency Calibration Low Intensity Test .3 ms, .5 sec; .5 A, 200 ohm 20 Hz 30 Hz 40 Hz Serial Number I Independent Obtained Values Voltage + Voltage - Current + Current - Hz 50 Hz 60 Hz 70 Hz 80 Hz 90 Hz 100 Hz 110 Hz 120 Hz SPECTRUM Reported Values Current Dynamic Imp Energy Charge Current Independent Values Dynamic Imp I 20 Hz L 30 Hz I 40 Hz 5 0 Hz 60 Hz 70 Hz 80 90 LIZ 100 Hz 110 Hz 120 Hz Charge (Page 10 of 15) ECT Device Test and Calibration Form Date Device Type Pulse Frequency Calibration High Intensity Test 1.5ms, 4 sec; .8A, 300 ohm I Serial Number Independent Obtained Values Voltage + Voltage - Current + Current - Hz 20 Hz 30 Hz 40 Hz 50 Hz 60 Hz 70 Hz 80 Hz 90 Hz 100 Hz 110 Hz 120 Hz 20 Hz 30 Hz 40 Hz 50 Hz 60 Hz 70 Hz 8O Hz 90Hz I 100 Hz 11O Hz I 1 120 Hz SPECTRUM Reported Values Independent Values Current Dynamic Energy Charge Current Dynamic Energy Imp Imp (-.I Charge (Page 11 of 15) ECT Device Test and Calibration Form Date Device Type Serial Number Current Intensity Calibration Low Intensity Test: .3 PW, 20 Hz, .5 sec; 200 ohm 0 A .600 A .700 A .800 A .900 A_ .500 A .600 A .700 A .800 A .900 A High Intensity Test 1.5 PW, 120 Hz, 4 sec, 300 ohm .500 A .600 A .700 A .800 A .900 A .500 A .600 A .700 A .800 A .900 A Independent Obtained Values Voltage + Voltage - Current + Current - Hz SPECTRUM Reported Values Independent Values Current Dynamic Energy Charge Current Dynamic Energy Imp Imp L. I Charge (Page 12 of 15) ECT Device Test and Calibration Form Date I Device Type I I Serial Number] Train Duration Calibration Low Settings .3 PW, 20 Hz, .8 A, 200 ohm 0.50 sec 0.75 sec 1 .00 sec 1.25 sec 1.50 sec 1.75 sec 2.00 sec 2.50 sec 3.00 sec 4 Independent Obtained Values Voltage + Voltage - Current + Current - Cur Duration Hz 0.50 sec 0.75 sec 1.00 sec 2.00 sec 2.50 sec 3.00 sec 4.00 sec SPECTRUM Reported Values Current Dynamic Energy Charge L. I Independent Values Current Dynamic Energy Charge Imp (Page 13 of 15) I ECT Device Test and Calibration Form Date Device Type Serial Number I Train Duration Calibration High Settings 1.5 PW, 120 Hz, .8 A, 300 ohm 0.5 sec 0.75 sec 1.00 sec 1.25 sec 1.75 sec 2.00 sec 2.50 sec 3.00 4.00 sec .00 sec 0.50 sec 0.75 sec 1.00 sec 1.25 sec 1.50 sec 1.75 sec 2.00 sec 2.50 sec Independent Obtained Values Voltage + Voltage - Current + Current - Cur Duration Hz 3.00 sec 4.00 sec SPECTRUM Reported Values Independent Values Current Dynamic Energy Charge Current Dynamic Energy Charge IMP 9 Imp (Page 14 of 15) ECT Device Test and Calibration Form Date 9 Device Type I Serial Number 150 ohm] 200 ohm 250 ohm 300 ohm] 350 ohm 400 ohm] 450 ohm] 500 ohm] 550 ohm] 150 ohm] 200 250 ohm] 1 300 ohm 350 ohm] 400 ohm] 450 ohm 500 ohm 550 ohm] Dynamic Impedance Calibration: Low Intensity 20 Hz, .3 msec, .5 sec, .8 A SPECTRUM Reported Values Independent Values Current Dynamic Energy Charge Current Dynamic Energy Charge Imp Imp Dynamic Impedance Calibration: High Intensity 120 Hz, 1.5 msec, 4 sec, .8 A SPECTRUM Reported Values Independent Values Current Dynamic Energy Charge Current Dynamic Energy Charge Imp Imp (Page 15 of 15) |
| home |